Publications

Polymer/enzyme-modified HF-etched carbon nanoelectrodes for single-cell analysis

Marquitan, M. and Ruff, A. and Bramini, M. and Herlitze, S. and Mark, M.D. and Schuhmann, W.

BIOELECTROCHEMISTRY
Volume: 133 Pages:
DOI: 10.1016/j.bioelechem.2020.107487
Published: 2020

Abstract
Carbon-based nanoelectrodes fabricated by means of pyrolysis of an alkane precursor gas purged through a glass capillary and subsequently etched with HF were modified with redox polymer/enzyme films for the detection of glucose at the single-cell level. Glucose oxidase (GOx) was immobilized and electrically wired by means of an Os-complex-modified redox polymer in a sequential dip coating process. For the synthesis of the redox polymer matrix, a poly(1-vinylimidazole-co-acrylamide)-based backbone was used that was first modified with the electron transfer mediator [Os(bpy)2Cl]+ (bpy = 2,2′-bipyridine) followed by the conversion of the amide groups within the acrylamide monomer into hydrazide groups in a polymer-analogue reaction. The hydrazide groups react readily with bifunctional epoxide-based crosslinkers ensuring high film stability. Insertion of the nanometre-sized polymer/enzyme modified electrodes into adherently growing single NG108-15 cells resulted in a positive current response correlating with the intracellular glucose concentration. Moreover, the nanosensors showed a stable current output without significant loss in performance after intracellular measurements. © 2020

« back